Description
Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Published: 2019-03-25
Score: 9.8 Critical
EPSS: < 1% Very Low
KEV: No
Impact: n/a
Action: n/a
AI Analysis

Analysis and contextual insights are available on OpenCVE Cloud.

Remediation

No vendor fix or workaround currently provided.

Additional remediation guidance may be available on OpenCVE Cloud.

Tracking

Sign in to view the affected projects.

Advisories
Source ID Title
EUVD EUVD EUVD-2015-3984 Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
History

No history.

Subscriptions

Pifzer Plum A\+3 Infusion System Plum A\+3 Infusion System Firmware Plum A\+ Infusion System Plum A\+ Infusion System Firmware Symbiq Infusion System Symbiq Infusion System Firmware
cve-icon MITRE

Status: PUBLISHED

Assigner: icscert

Published:

Updated: 2024-08-06T06:04:00.957Z

Reserved: 2015-05-12T00:00:00.000Z

Link: CVE-2015-3953

cve-icon Vulnrichment

No data.

cve-icon NVD

Status : Modified

Published: 2019-03-25T17:29:00.623

Modified: 2024-11-21T02:30:08.053

Link: CVE-2015-3953

cve-icon Redhat

No data.

cve-icon OpenCVE Enrichment

No data.

Weaknesses