* ELI 380 Resting Electrocardiograph:
Versions 2.6.0 and prior;
* ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph:
Versions 2.3.1 and prior;
* ELI 250c/BUR 250c Resting Electrocardiograph: Versions 2.1.2 and prior;
* ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph:
Versions 2.2.0 and prior.
Analysis and contextual insights are available on OpenCVE Cloud.
Vendor Solution
Hillrom has released software updates for all impacted devices to address these vulnerabilities. New product versions that mitigate these vulnerabilities are available as follows: * Welch Allyn ELI 380 Resting Electrocardiograph: available Q4 2023 * Welch Allyn ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph: available May 2022 * Welch Allyn ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph: available Q4 2023 Hillrom recommends users upgrade to the latest product versions. Information on how to update these products can be found on the Hillrom disclosure page https://hillrom.com/en/responsible-disclosures/ .
Vendor Workaround
Hillrom recommends the following workarounds to help reduce risk: * Apply proper network and physical security controls. * Ensure a unique encryption key is configured for ELI Link and Cardiograph. * Where possible, use a firewall to prevent communication on Port 21 FTP service, Port 22 SSH (Secure Shell Connection), and Port 23 Telnet service.
Tracking
Sign in to view the affected projects.
| Source | ID | Title |
|---|---|---|
EUVD |
EUVD-2022-30948 | An improper access control vulnerability may allow privilege escalation.This issue affects: * ELI 380 Resting Electrocardiograph: Versions 2.6.0 and prior; * ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph: Versions 2.3.1 and prior; * ELI 250c/BUR 250c Resting Electrocardiograph: Versions 2.1.2 and prior; * ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph: Versions 2.2.0 and prior. |
Fri, 07 Feb 2025 19:15:00 +0000
| Type | Values Removed | Values Added |
|---|---|---|
| Metrics |
ssvc
|
Fri, 07 Feb 2025 17:15:00 +0000
| Type | Values Removed | Values Added |
|---|---|---|
| Description | An improper access control vulnerability may allow privilege escalation.This issue affects: * ELI 380 Resting Electrocardiograph: Versions 2.6.0 and prior; * ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph: Versions 2.3.1 and prior; * ELI 250c/BUR 250c Resting Electrocardiograph: Versions 2.1.2 and prior; * ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph: Versions 2.2.0 and prior. | |
| Title | Improper Access Control Vulnerability in ELI Electrocardiograph Devices | |
| Weaknesses | CWE-284 | |
| References |
| |
| Metrics |
cvssV3_1
|
Subscriptions
No data.
Status: PUBLISHED
Assigner: Baxter
Published:
Updated: 2025-02-07T18:49:22.506Z
Reserved: 2022-03-03T22:04:24.534Z
Link: CVE-2022-26389
Updated: 2025-02-07T18:48:32.460Z
Status : Deferred
Published: 2025-02-07T17:15:22.130
Modified: 2026-04-15T00:35:42.020
Link: CVE-2022-26389
No data.
OpenCVE Enrichment
No data.
EUVD