Description
In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text
patient data to a hard-coded public IP address when a patient is hooked
up to the monitor. This could lead to a leakage of confidential patient
data to any device with that IP address or an attacker in a
machine-in-the-middle scenario.
Published: 2025-01-30
Score: 8.2 High
EPSS: < 1% Very Low
KEV: No
Impact: n/a
Action: n/a
AI Analysis

Analysis and contextual insights are available on OpenCVE Cloud.

Remediation

Vendor Solution

Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks. Please note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA's safety communication https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication .

Tracking

Sign in to view the affected projects.

Advisories
Source ID Title
EUVD EUVD EUVD-2025-1815 In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an attacker in a machine-in-the-middle scenario.
History

Wed, 12 Feb 2025 21:15:00 +0000

Type Values Removed Values Added
Metrics ssvc

{'options': {'Automatable': 'no', 'Exploitation': 'none', 'Technical Impact': 'partial'}, 'version': '2.0.3'}


Fri, 31 Jan 2025 17:30:00 +0000

Type Values Removed Values Added
Description In its default configuration, the affected product transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an attacker in a machine-in-the-middle scenario. In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an attacker in a machine-in-the-middle scenario.
Metrics cvssV3_1

{'score': 5.9, 'vector': 'CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N'}


Fri, 31 Jan 2025 16:45:00 +0000


Thu, 30 Jan 2025 18:30:00 +0000

Type Values Removed Values Added
Description In its default configuration, the affected product transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an attacker in a machine-in-the-middle scenario.
Title Exposure of Private Personal Information to an Unauthorized Actor vulnerability in Contec Health CMS8000 Patient Monitor
Weaknesses CWE-359
References
Metrics cvssV4_0

{'score': 8.2, 'vector': 'CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N'}


Subscriptions

No data.

cve-icon MITRE

Status: PUBLISHED

Assigner: icscert

Published:

Updated: 2025-02-12T20:41:36.789Z

Reserved: 2025-01-23T18:11:20.770Z

Link: CVE-2025-0683

cve-icon Vulnrichment

Updated: 2025-01-31T15:35:10.389Z

cve-icon NVD

Status : Deferred

Published: 2025-01-30T19:15:14.460

Modified: 2026-04-15T00:35:42.020

Link: CVE-2025-0683

cve-icon Redhat

No data.

cve-icon OpenCVE Enrichment

No data.

Weaknesses